The E-ELT secondary mirror support spiders emit infrared radiation that contaminates HARMONI observations. The Pupil Mask Rotator keeps a thermal mask aligned with those spiders at all times, compensating for the telescope's field rotation.
COTS component selection
The starting constraint was to use COTS components adapted to cryogenic conditions. The Applimotion UTS-89-A-25-A hollow-shaft motor was chosen (SmCo magnets, more thermally stable than neodymium at low temperatures) along with the Netzer DS-90 capacitive encoder (19-bit, ±36 arcsec resolution).
Cryogenic encoder characterisation
With no cryogenic reference encoder available, a contactless angular measurement system was developed using machine vision: an IAC-built cryogenic camera photographs two circular marks on the shaft, and a centroid algorithm in Python determines the angle to better than 50 arcsec accuracy.
Control system
The position loop is closed with TwinCAT 3 running on a Beckhoff PLC. A linear current driver — not PWM — minimises electromagnetic noise inside the optical instrument. The achieved position error was below 100 arcsec, meeting the project specification.