HARMONI's cryogenic mechanisms had to be characterised under representative conditions before integration into the instrument. The test bench was the first step: a system that could reproduce the cryo-vacuum environment in a controlled and repeatable way.
Design requirements
The bench had to reach temperatures below 80 K via indirect liquid nitrogen immersion, maintain a working pressure of 10⁻⁵ mbar, allow electrical connections for motors and encoders from outside without breaking the vacuum, and facilitate mechanism mounting and dismounting under test.
Adopted solution
A double-wall stainless steel cryostat with vacuum insulation between walls was designed. Electrical connections were solved using commercially certified vacuum feedthroughs. The mechanism support is modular: a threaded aluminium base plate that accepts different test configurations.
Results
The bench served for all cryogenic tests of both the Pupil Mask Rotator and the Voice Coil. It reached 25 K minimum temperature at the mechanism platform. Electrical connections maintained hermeticity throughout the entire test campaign.