Test Bench for Cryogenic Mechanisms
Cryogenic test bench at the IAC laboratory. Liquid-nitrogen cryostat with electrical feedthroughs for motor and encoder control.

HARMONI's cryogenic mechanisms had to be characterised under representative conditions before integration into the instrument. The test bench was the first step: a system that could reproduce the cryo-vacuum environment in a controlled and repeatable way.

Design requirements

The bench had to reach temperatures below 80 K via indirect liquid nitrogen immersion, maintain a working pressure of 10⁻⁵ mbar, allow electrical connections for motors and encoders from outside without breaking the vacuum, and facilitate mechanism mounting and dismounting under test.

Adopted solution

A double-wall stainless steel cryostat with vacuum insulation between walls was designed. Electrical connections were solved using commercially certified vacuum feedthroughs. The mechanism support is modular: a threaded aluminium base plate that accepts different test configurations.

The greatest challenge was not the cold, but ensuring that the electrical connections remained hermetic at 10⁻⁵ mbar after hundreds of thermal cycles.

Results

The bench served for all cryogenic tests of both the Pupil Mask Rotator and the Voice Coil. It reached 25 K minimum temperature at the mechanism platform. Electrical connections maintained hermeticity throughout the entire test campaign.

End of project